gas terminal type H
Description
The gas terminal type H was developed in the years between 1996 to 1999 in view of the forthcomming european standard. Since 1999 they are in use all over the world. It is a gas terminal for medical gases with the DIN 13260-2 connector coding (other connector codings or types of gases upon request). An optional sintered filters protects the subsequent seals, connection fittings and patients against pollution from the supply system. While disengage the connection armatures this is held in park position compulsorily to prevent a drop out of the valve.
specifications
classification according to directive 93/42/EWG | class IIa |
operating pressure | compressed gas: 400 kPa to 500 kPa ± 20 % vacuum: -100 kPa to -40 kPa |
dimension with cover plate (WxHxD), in-wall | 89x89x10 mm |
dimension with cover plate (WxHxD), on-wall | 82x82x62 mm |
different installation situations
ceiling supply units
REF-number | description |
---|---|
5752 0218 | basic block H (hose), pressure gases LW8 |
5752 4331 | basic block H (hose), vacuum LW10 |
5743 1335-1339 | fuse box O2, AIR, N2O, VAC, CO2 |
5750 1657 | locking part EN, Type H (not required for VAC) |
5750 3093-3097 | plug receptacle H FN O2, AIR, N2O, VAC, CO2 |
5752 2377-2381 | plug receptacle H ISO O2, AIR, N2O, VAC, CO2 |
optional | |
5750 1706 | sintered metal filter (not for VAC) |
installation channel and wall supply units
REF-number | description |
---|---|
5750 0049 | basic block H (copper pipe) |
5743 1958 | mounting kit |
5750 3093-3097 | socket H FN O2, AIR, N2O, VAC, CO2 |
5752 2377-2381 | socket H ISO O2, AIR, N2O, VAC, CO2 |
5743 1335-1339 | fuse box O2, AIR, N2O, VAC, CO2 |
5750 1657 | locking part EN, Type H (not required for VAC) |
5752 0246 | mounting plate |
5731 1152 | pop rivet |
optional | |
5752 0492 | cover plates holder H-valve |
5752 0487 | cover plate H-valve |
5750 4645 | circlip |
5743 0158 | Ejot-Screw |
5750 1706 | sintered metal filter (not for VAC) |
massive and lightweight walls
REF-number | description |
---|---|
5750 0270 | installation housing EN (in-wall housing) |
5750 3093-3097 | socket H FN O2, AIR, N2O, VAC, CO2 |
5752 2377-2381 | socket H ISO O2, AIR, N2O, VAC, CO2 |
5750 1657 | locking part EN, Type H (not required for VAC) |
5743 1335-1339 | fuse box O2, AIR, N2O, VAC, CO2 |
5752 0487 | cover plate H-valve |
5743 0161 | zinc cast dowe |
5743 0162 | screw |
5743 0158 | Ejot-Screw |
optional | |
5750 1706 | sintered metal filter (not for VAC) |
on-wall mounting
REF-number | description |
---|---|
5752 0951-0954, 9000 1048 | wall mounted outlet type H FN O2, AIR, N2O, VAC, CO2 |
5752 4187-4191 | Wall mounted outlet type H ISO O2, AIR, N2O, VAC, CO2 |
documents
operating instructions (PDF)
installation instructions (PDF)
maintenance manual (PDF)